Information Messages (Displayed by the i Icon)

List of Messages

  • Export platforms have identical names.
  • Export to [export destination] failed due to insufficient access rights.
  • Export to [export destination] failed due to insufficient disk space.
  • Export to [export destination] failed.
  • Treatment Plan was separated by Fraction Groups and more than one RT Plan was exported.
  • Detected regions of the couch top that do not intersect the SliceSet. Nonintersecting regions will not be exported.
  • Contour points were reduced
  • Structure [object name] was changed for export as it was reconstructed on the reference image data set.
  • Structure [object name] could not be exported as it was reconstructed on the reference image data set and appears to be too small.
  • Beam Dose was exported, but one or more beams are split. Thus, the RT Dose object now references more than one beam.
  • The RT Dose Volume object does not contain a plan reference, as the plan was separated and dose export for the entire plan was used.
  • [clinical indication] [protocol name] is not available. Name is used in multiple files: [file 1], [file 2].
  • Image Modality (XT) not supported
  • DICOM image(s) of modality XT cannot be processed by the current version of the application.
  • An error occurred while loading.
  • Unsupported data found.
  • Unsupported data was found during loading. Review the patient data carefully for completeness.
  • Plan not valid for this application.
  • This plan has an invalid format.
  • The plan to load has a format that cannot be interpreted.
  • An error occurred while loading.
  • Multiple plans selected.
  • More than one plan is selected for loading. Go back and select only a single plan.
  • Continuing the optimization of a plan created with a version of the application other than the current one is not supported.
  • Continuing the optimization of a plan created with a previous version of the application is not supported.
  • Detected a problem with the linac constraints. The following problem(s) occurred:
  • The PTV is too big and therefore cannot be irradiated sufficiently with the current MLC and/or fluence mode.
  • Dose in PTV, PTV-Boost, or Boost cannot be calculated. This can e.g. happen if parts of the PTV are located outside the outer contour, or if the Boost is identical to the PTV.
  • The External Outer Contour [contour name] has been removed.
  • The External Outer Contour [contour name] can not be used. It has been removed.

  • Multiple External Outer Contours were found. There must be exactly one External Outer Contour. Go back to the selected data and check this.

  • Object [object name] is not fused to any image set. Go back to Smartbrush and Image Fusion and check this.

  • Empty object found.
  • The object [object name] is empty. Outline the object correctly or remove it from the selection.

  • Empty objects found.
  • The following objects are empty: [object names] Outline the objects correctly or remove them from the selection.

  • The image set of Object [object name] is not loaded. Go back to the selected data and check this.

  • Multiple objects are named identically.
  • Found more than one object with name [object name]. The objects’ names must be unique. Go back to the selected data and check this.

  • The following objects have identical names: [object names] The objects’ names must be unique. Go back to the selected data and check this.

  • The planning CT is deformed. It must not be deformed. Go back to the selected data and check this.

  • The planning CT is localized. It must not be localized. Go back to the selected data and check this.

  • The planning image set is not of Modality CT. It must be a CT image set. Go back to the selected data and check this.

  • The planning CT is not a supine image. Only supine images are supported. Go back to the selected data and check this.

  • The planning CT is not a head first image set. Only head first images are supported. Go back to the Selected Data and check this.

  • The planning CT is a Cone Beam CT. This could lead to dose calculations being significantly off and is therefore not allowed. Go back to the selected data and check this.

  • The planning CT contains tilted image slices. Tilted image slices are not allowed.

    Go back to the selected data and check this.

  • The planning CT has too many slices.
  • The planning CT has [number of slices] slices. A maximum number of [number of slices] slices is supported. Go back to the selected data and check this.

  • Multiple image sets with Modality CT were found. There must be exactly one image dataset with Modality CT. Go back to the selected data and check this.

  • No image set with Modality CT was found. There must be exactly one image dataset with Modality CT. Go back to the selected data and check this.

  • Not all image sets are fused. Go back to the selected data and check this.
  • Object [object name] is set to invisible.
  • Volume of object(s) too small.

  • The following objects were found to be too small: [object names]. Only objects with volume bigger than [volume] mm³ are allowed. Go back to the selected data and check this.
  • There were unspecified problems during import.
  • Could not save RT Plan. Contact Brainlab support.
  • The loaded plan was created with a previous software version. Optimization, saving and export are disabled. To start a new optimization, select a different clinical protocol.

  • The license for this application has expired. Optimization, saving and export are disabled.

  • The license for this application has expired but no plan was loaded. The application cannot continue.
  • No more than one Most Important OAR is allowed. Currently there are:
  • Arc Duplication added [number of arcs] arc(s), therefore the total number of arcs is [number of arcs]. Manual modification of arcs is not possible.
  • Failed to calculate the tissue model.
  • No PTV was found. Go back to the selected data and check this.
  • Boost is partly outside of PTV.
  • The boost object [boost name] is outlined partly outside of the PTV object [PTV name]. The boost object must be fully enclosed within the PTV. Go back to the selected data and check this.
  • Multiple PTVs found.
  • The following PTVs were found: [PTV names] Only one PTV and possibly a single enclosed boost are allowed. Go back to the selected data and check this.
  • The treatment target consists of disconnected volumes. This is not supported by Spine SRS. To treat this target, combine the volumes into a single target volume.
  • Volume of some Object(s) too small.
  • The following Objects were found to be too small: [object names] Only Objects with volume bigger than 0.01cm³ are allowed. Go back to the selected data and check this.
  • Boost volume too big.
  • The volume of the boost object [boost name] has to be smaller than the volume of the PTV object [PTV name] by at least 0.01cm³. Please go back to the selected data and check this.
  • Dose calculation failed. Contact Brainlab support.
  • Dose for the plan is being calculated using PencilBeam. Due to the small size of the PTVs the plan likely defines a considerable amount of small fields (of 10 mm equiv. square fields size or smaller). It is recommended to use Monte Carlo instead for dose calculation of such small fields in order to have decent accuracy of calculation results.
  • Dose calculation is based on special set: Contrast CT.