Cranial IGS System Ultrasound Navigation

Содержание

How to Set Up the Camera

Step

1.

Adjust the camera to the required position.

Position the camera approximately 1.2 – 1.8 meters (4 – 6 feet) away from the surgical field.

2.

Plug in and turn on the system. Refer to the platform’s System and Technical User Guide (e.g., Kick, Curve).

3.

Start the software.

4.

Ensure the reference array is visible to the camera.

The following view is displayed to indicate whether the array is visible to the camera.



If the array is not visible, adjust the camera distance and angle until the array representation is visible.