Messages in the Printout (Traffic Lights)

List of Messages

TypeText
InfoDose calculations have not started. Activate dose display to enable dose distribution related plan status checks.
Info

Dose calculation is based on special set: [modality name].

ErrorLess than 2 arc control points are defined. Treatment is not possible. Verify the arc start and stop angles.
WarningLess than 3 arc control points are defined. Treatment may result in a controller error. Verify the arc start and stop angles.
ErrorThe MU per degree value is lower than defined in the machine profile.
ErrorThe MU per degree value is higher than defined in the machine profile.
ErrorThe gantry rotation range is not valid. Verify the arc start and stop angles.
ErrorDanger of table collision. Please verify.
WarningRisk of table collision. Please verify.
ErrorThe table angle is invalid. Treatment is not possible.
ErrorThe ring rotation angle is not valid. Treatment is not possible.
ErrorThe collimator angle is not valid. Treatment not possible.
WarningThe arc defined by start and stop angles is ambiguous. Please verify that the desired sector is treated.
ErrorThe gantry rotation range is not valid. Verify the arc start and stop angles.
Error

The time to deliver the desired dose ([time] s) exceeds the machine limit ([time] s). Please verify.

Error

The MU ([number of units] MU) value of one of this beam’s/arc’s export portions is above the allowed MU ([number of units] MU) configured for the machine. Contact Brainlab support.

WarningThe number of portions into which this element is split exceeds 10. Please verify.
Info

The element is split into [number of portions] portions because the MU value ([number of units] MU) is larger than the maximum allowed MU ([number of units] MU).

Error

MU ([number of units] MU) is larger than the maximum allowed MU ([number of units] MU). (Note: splitting to several portions is not supported by the application for this treatment). Contact Brainlab support.

WarningThe dose needs to be recalculated because the loaded plan was saved without dose. The dose might deviate slightly from the original plan.
ErrorThe name of the beam or arc contains invalid characters.
ErrorThe name of the beam or arc is empty.
ErrorThe name of the beam or arc contains either of the characters ‘(‘ or ‘)’.
Warning

The data set used as Alignment Set has been scanned [scan orientation], while the orientation for treatment is configured [scan orientation]. Contact Brainlab support.

Warning

The data set used as Reference Set has been scanned [scan orientation], while the orientation for treatment is configured [scan orientation]. Contact Brainlab support.

ErrorThe treatment orientation of the patient is undefined. This status is critical! Please consult Brainlab support and do not use the current plan for treatment.
ErrorThe image and patient orientation for treatment of the current plan have not been acknowledged. This status is critical. Please consult Brainlab support and do not use the current plan for treatment.
Warning

The plan references the following empty object(s) as PTV or OAR: [object names]. These objects should be correctly segmented if required.

WarningThe Dose Calculations are based on a Reference Set of type Cone Beam and may be significantly off. It is highly recommended to perform an update CT scan that is used as Reference Set.
WarningHeterogeneity correction is turned off. The table top added to the tissue model is therefore assumed to be water-equivalent. Contact Brainlab support.
InfoHeterogeneity correction is turned off. The area inside the detected patient contour is therefore assumed to be water-equivalent. Contact Brainlab support.
Warning

The plan intent is set to [plan intent value].

Error

Encountered [number of duplicates] beams or arcs whose names are not unique. Beam- or arc names involved are: [duplicate names]. Contact Brainlab support. (Note: Name comparison is case-insensitive).

WarningThe dose resolution for the current plan exceeds the limit of 10 mm. Contact Brainlab support.
ErrorThe Alignment Set must be the same Slice Set as the Reference Set, however two different Slice Sets are used in this Treatment Plan. If you continue, in worst case, we expect geometric inaccuracies when you export the plan and position the patient for treatment. Please contact Brainlab support.
InfoSome values of the selected HU to ED conversion table are outside plausible ranges.
WarningThe selected HU to ED conversion table is for acceptance only and must not be used for any patient treatment.
WarningThe selected HU to ED conversion table is not approved and must not be used for any patient treatment.
WarningThe machine profile was not validated by Brainlab. Treatment is not possible.
WarningThe active machine profile loaded from the current plan is not contained in the list of installed machine profiles. The profile may be outdated or from an external source. Verify.
ErrorThe isocenter coordinate is not inside the patient’s tissue.
InfoUnsupported dose rate. Contact Brainlab support.
WarningThe active machine profile is not approved.
WarningThe active machine profile and its beam data are not approved.
Warning 
WarningThe active machine profile is for acceptance only and must not be used for patient treatment.
Warning

The isocenter coordinate [coordinate name] is not assigned the prefix [coordinate prefix]. Contact Brainlab support.

Warning

The treatment group [group name] is not assigned the prefix [group prefix]. Contact Brainlab support.

Error

The isocenter coordinate name [coordinate name] is either not unique or too long. Contact Brainlab support.

Error

The treatment group name [group name] is either not unique or too long. Contact Brainlab support.

Warning

The treatment group [group name] uses the identifier [identifier name]. This is also used by a non-related isocenter coordinate. Contact Brainlab support.

Warning

The treatment group [group name] and its isocenter coordinate [coordinate name] are each identified by a different number. The same number should be used where possible. If several treatment groups are assigned to the same isocenter coordinate, unique alphanumeric identifiers should be used. Contact Brainlab support.

InfoElement order and gantry rotation direction might be improved for treatment.
InfoThe plan contains closed fields (MLC).
InfoThe isocenter is blocked (MLC). Please verify.
InfoThe plan contains closed fields (jaws).
InfoThe isocenter is blocked (jaws). Please verify.
WarningThe jaw positions expose the MLC carriage box. Contact Brainlab support.
WarningThe jaws are overlapping the MLC field. Please verify.
WarningThe planned leaf gap of at least one closed leaf pair is not covered by the jaws. Please check your plan carefully.
InfoThe planned leaf gap of at least one closed leaf pair is within a 2 mm range of the jaw edge and might not be covered by the jaws during delivery. Please check your plan carefully.
Warning

The maximum beamlet size is smaller than the [grid resolution value]. Dose calculation accuracy may be limited.

Warning
For a considerable number of fields of the arc:

  • The minimum field extension is smaller than four times the [grid resolution value]. Dose calculation accuracy may be limited.
  • The minimum field extension is smaller than four times the dose grid resolution. Dose calculation accuracy may be limited.
InfoApplication is in RESEARCH mode. Export is not possible.
InfoThis treatment plan has image sets containing Standardized Uptake Values (SUV). The displayed SUV can vary depending on the manufacturer of the PET scanner. Always compare the displayed values with the SUV obtained directly at the scanner before using them.
WarningThis data must only be used for commissioning and plan comparison and not for creating clinical treatment plans. Export is not possible.
Error

Gantry lower angle [gantry lower angle]° of Arc Plane [arc name] is greater than gantry upper angle [gantry upper angle]°.

Error

Gantry angles [gantry lower angle]° – [gantry upper angle]° of Arc Plane [arc name] do not have the required minimal arc span of [arc span]°.

Error

Table angle [table angle]° of Arc Plane [arc name] is incompatible with the selected machine profile.

Error

Gantry angles [gantry lower angle]° – [gantry upper angle]° of Arc Plane [arc name] are incompatible with the selected machine profile.

Error

Duplicate arc plane definitions for the table angle(s) [table angles].

ErrorCoplanar arc plane definitions for the table angle(s) [table anges].
Error

Collimator angle [collimator angle]° of Arc Plane [arc name] is incompatible with the selected machine profile.

Error

The boost’s Tolerated Coverage Volume [volume]% is greater than the Desired Coverage Volume [volume]%.

Error

The boost’s maximum dose constraint [dose] Gy is less than the prescription dose [dose] Gy.

Error

The boost’s prescription dose constraint [boost reference dose] Gy is less than the target’s prescription dose [ptv reference dose] Gy.

ErrorSRS prescription is selected for the boost, but no maximum dose constraint is set.
 IDL prescription [current IDL]% is outside of allowed range of [minimum IDL]% – [maximum IDL]%.
ErrorThe clinical protocol has unsupported version $version.
Info

[clinical indication] [protocol name] is from a previous application version: [protocol type].

Error

The most important OAR [oar name] is not part of the OAR Constraints list.

Error

Constraint # [number] of OAR [oar name]: relative volume is out of range: [volume]%.

Warning

The current treatment plan contains objects that are either not mentioned in the protocol or are being ignored because they have the wrong role/type. These objects will not be taken into account during optimization.

Names of ignored objects:

Warning

The protocol contains object types that are not available for the current treatment plan. These object types will not be taken into account during optimization.

Missing types:

Warning

The following dose lines are above the prescription dose of [dose] Gy and will not be displayed: [hidden dose lines].

Error

The Tolerated Coverage Volume [volume]% is greater than the Desired Coverage Volume [volume]%.

Error

The maximum prescribed dose constraint [dose] Gy is less than the prescription dose [dose] Gy.

ErrorSRS prescription is selected, but no maximum dose constraint is set.
WarningThe selected clinical protocol is for Machine QA only. Not for clinical use.
WarningThe selected clinical protocol is provided for Machine QA demo purposes only. Not for clinical use.
WarningThe selected clinical protocol is provided for demo purposes only. Not for clinical use.
WarningVMAT deliverability warning:
Warning

The PTV overlaps with the following OARs: [oar names]. This can lead to poor plan results.

Warning
For a considerable number of fields of the arc, the equivalent field size is outside the range of measured field sizes:

  • it is smaller than the depth dose field size range

  • it is larger than the depth dose field size range

  • it is smaller than the scatter jaw-field size range

  • it is larger than the scatter jaw-field size range

  • it is larger than the scatter MLC-field size rangeit is smaller than the scatter MLC-field size range

Warning

it is smaller than the scatter MLC-field sizeInsufficient scatter table data for jaw field size [field size] mm: the value for MLC field size [field size] mm should be measured and not be copied. See the Technical Reference Guide Brainlab Physics for more details.