Information Messages (Displayed by the i Icon)

List of Messages

  • Treatment plan was separated by fraction groups and more than one RT Plan was exported.
  • Detected regions of the couch top that do not intersect the slice set. Nonintersecting regions will not be exported.
  • Contour points were reduced.
  • Structure [object name] was changed for export as it was reconstructed on the reference image data set.
  • Structure [object name] could not be exported as it was reconstructed on the reference image data set and appears to be too small.
  • Beam Dose was exported, but one or more beams are split. Thus, the RT Dose object now references more than one beam.
  • The RT Dose Volume object does not contain a plan reference, as the plan was separated and dose export for the entire plan was used.
  • Image modality (XT) not supported.
  • DICOM image(s) of modality XT cannot be processed by the current version of the application.
  • An error occurred while loading.
  • Unsupported data found.
  • Unsupported data was found during loading. Review the patient data carefully for completeness.
  • Plan not valid for this application.
  • This plan has an invalid format.
  • The format of the plan to load cannot be interpreted.
  • An error occurred while loading.
  • Multiple plans selected.
  • More than one plan is selected for loading. Go back and select only a single plan.
  • The external outer Contour [contour name] has been removed.
  • The external outer Contour [contour name] can not be used. It has been removed.
  • Multiple external outer Contours were found. There must be exactly one external outer Contour. Go back to the selected data and check this.
  • Object [object name] is not fused to any image set. Go back to Smartbrush and Image Fusion and check this.
  • Empty object found.
  • Empty objects found.
  • The object [object name] is empty. Outline the object correctly or remove it from the selection.
  • The following objects are empty: [object names]. Outline the objects correctly or remove them from the selection.
  • The image set of object [object name] is not loaded. Go back to the selected data and check this.
  • Multiple objects are named identically.
  • Found more than one object with the name [object name]. The objects’ names must be unique. Go back to the selected data and check this.
  • The following objects have identical names: [object names] The objects’ names must be unique. Go back to the selected data and check this.
  • The planning CT is deformed. It must not be deformed. Go back to the selected data and check this.
  • The planning CT is localized. It must not be localized. Go back to the selected data and check this.
  • The planning image set is not of modality CT. It must be a CT image set. Go back to the selected data and check this.
  • The planning CT is not a supine image. Only supine images are supported. Go back to the selected data and check this.
  • The planning CT is not a head first image set. Only head first images are supported. Go back to the selected data and check this.
  • The planning CT is a Cone Beam CT. This could lead to dose calculations being significantly off and is therefore not allowed. Go back to the selected data and check this.
  • The planning CT contains tilted image slices. Tilted image slices are not allowed. Go back to the selected data and check this.
  • The planning CT has [number] slices. A maximum number of [number] slices is supported. Go back to the selected data and check this.
  • Multiple image sets with modality CT were found. There must be exactly one image dataset with modality CT. Go back to the selected data and check this.
  • No image set with modality CT was found. There must be exactly one image dataset with modality CT. Go back to the selected data and check this.
  • Not all image sets are fused. Go back to the selected data and check this.
  • Object [object name] is set to invisible.
  • Volume of object(s) too small.
  • The following objects were found to be too small: [object names] Only objects with volume bigger than [volume] mm³ are allowed. Go back to the selected data and check this.
  • There were unspecified problems during import.
  • Could not save plan. Contact Brainlab support.
  • Loaded plan from an older version.
  • The loaded plan was created with a previous software version. Optimization, saving and export are disabled. To start a new optimization, select a different clinical protocol.
  • The license for this application has expired. Optimization, saving and export are disabled.
  • The license for this application has expired but no plan was loaded. The application cannot continue.
  • Failed to calculate the tissue model. Check the logfile for details.
  • The dose was recalculated.
  • The dose was recalculated at startup because the loaded plan was saved using an older software version. The dose might deviate slightly from the original plan.
  • Multiple images set were found. There must be exactly one image dataset with modality CT. Go back to the selected data and check this.
  • The data is not compatible with [application mode]: they were written by an unsupported application.
  • Compartment for Phantom Definition Plans [compartment ID] doesn’t seem to be configured at connected DicomProxy (or not at all configured upstream DicomProxies). Please have Brainlab Support check configuration of your DicomProxy.
  • Loading data without a plan is not supported in [application mode].
  • Transferring phantom data to treated patient failed. See logfile for details.
  • Dose for the plan is being calculated using PencilBeam. Due to the small size of the PTVs the plan likely defines a considerable amount of small fields (of 10 mm equiv. square fields size or smaller). It is recommended to use Monte Carlo instead for dose calculation of such small fields in order to have decent accuracy of calculation results.
  • Dose calculation is based on special set: Contrast CT.