Dose Review

Dose Option Template

VariableDescription

Type

Defines the type of protocol and software version.

Name

Defines the protocol name.

Purpose

Defines the purpose of the protocol.

Demo: Protocol is provided for demo purposes only. Not for clinical use.

Clinical Use: Protocol is suitable for clinical use.

Dose Display

Default Prescription Value

Defines the absolute dose value for normalization of the prescribed dose in Gy. This value overwrites the loaded normalization dose and cannot be changed in Dose Review if specified.

Optional

Dose Relation (%)

Defines the percentage for normalization of the prescribed dose. If this value is not specified only absolute dose evaluation is possible.

Highlight Dose (Gy)

Enter the default value in Gy for the free highlight field in the Dose Display function. This value must be between 1.0 – 99.9 Gy.

Show Half Prescription

Choose to enable or disable the half prescription value in the Dose Display function.

True: Enable the half prescription value.

False: Disable the half prescription value (default).

Show Maximum Limit

Choose to enable or disable the maximum dose limit.

True: Enable the maximum dose limit.

False: Disable the maximum dose limit.

Maximum Limit (%)

Enter the relative maximum dose limit. This value must be between 100 – 200%.

Show Absolute Dose

Choose to display absolute or relative dose values in the dose legend as the default. You can change this setting in the software by double-clicking on the dose legend.

True: Display the absolute dose.

False: Display the relative dose.

Dose Lines (Value List)

Dose Line 1…

You can define a maximum of three critical dose lines in the Dose Display function according to your needs.

Maximum: 3 values.

Dose (Gy)

Enter the dose (Gy) for each dose line. This value must be between 1.0 – 99.9 Gy.