xBrain Data (iPlan RT)

Loading Data

To load data created in iPlan RT Image and iPlan RT Dose, first convert the data to the DICOM Raw format through the xBrainConverter. This data should only be used for commissioning and plan comparisons and not for creating clinical treatment plans. Export is not possible.

xBrainConverter Restrictions

The following restriction apply to xBrain converted data:

  • External outer contours imported from xBrainConverter cannot be cropped. Adjusting the couch top has no effect on the outer contour.

  • xBrainConverter does not convert or load any dose or plan information or isocenters (e.g., beams, arcs).

  • xBrainConverter does not convert any data if the CT is localized.

Restricted Plan Intent

As soon as xBrain converted data is loaded, the Plan Intent is automatically set to Research. In research mode, the Export function is disabled and therefore clinical treatment is not possible. Imported xBrain converted data is for commissioning and comparing treatment plans created in this software with plans created by iPlan RT or other third-party treatment planning systems.

Object Definitions

The software only recognizes the following objects for irradiation:

  • iPlan RT Image: Under Object Creation, only Structure = Tumor (not PTV, GTV or CTV, etc) is recognized. If Structure is set to PTV in iPlan RT Image, the Elements assign this object as Type = PTV and Role = Undefined. Objects of Type = PTV are placed under Others and are not recognized as candidates for treatment even if the objects were previously treated in iPlan RT. You can redefine these objects in SmartBrush prior to starting this software.

  • iPlan RT Dose: Under Object Creation, only Structure = Tumor (not PTV, GTV or CTV, etc) is recognized. If Structure is set to PTV in iPlan RT Dose, the Elements assign this object as Type = PTV and Role = Undefined. Objects of Type = PTV are not recognized as candidates for treatment even if the objects were previously treated in iPlan RT. You can redefine these objects in SmartBrush prior to starting this software.

RT QA

Export is activated within the RT QA Element.